Evaluierungen nach ACVIM

Shih et al. 2010: Petmap vs. invasive

grimred Normotensive: Bias SAP - 14,7 mmHg, MAP - 16,4 mmHg, DAP - 14,1 mmHg
Hypotensive: Bias SAP - 32 mmHg, MAP - 24,2 mmGh, DAP - 16,8 mmHg

Evaluation Doppler and HDO vs. Telemetry in conscious cats (Martel et al, in preparation)

Below table 2.40 Summary of the results within an apparatus across the three technicians

Doppler: grimred Apparatus Regression coefficient Bias Limits of agreement Paired measurements (%) within ± 10 mmHg (1) Paired measurements (%) within ± 20 mmHg (2)
Doppler 1 0.951501 21.9 ± 23.5 15 47
Doppler 2 0.948206 16.2 ± 25 27 61
Doppler 3 0.951501 18.3 ± 26 25 52

Systolic arterial blood pressure was measured both by Doppler and Telemetry methods in the conscious cat.
(1) percentage of paired measurements lying within a difference less than 10 mmHg between the two methods.
(2) percentage of paired measurements lying within a difference less than 20 mmHg between the two methods.

Below table 2.1 Summary table

HDO hdosmiley Animal No. Regression coefficient Bias Limits of agreement Paired measurements (%) within ± 10 mmHg (1) Paired measurements (%) within ± 10 mmHg (2)
2008017 0.913311 -4.3 ± 23.9 97.0 97.0
2008018 0.938719 -2.7 ± 20.7 84.0 100.0
2008019 0.845708 -5.1 ± 31.9 87.0 93.0
2008020 0.946730 -3.2 ± 17.4 91.0 100.0
2008024 0.975803 1.6 ± 13.4 91.0 97.0
2008025 0.917352 0.7 ± 22.0 78.0 91.0
Mean 0.922937 -2.2 ± 21.6 88.0 96.3
SD 0.017980 1.1 2.6 2.7 1.5
N 6.000000 6.0 6.0 6.0 6.0

- 12 Punkte| Doppler | Ozillometrie | HDO | ACVIM consensus | Evaluierung nach ACVIM